    VACUMED 代谢模拟器METABOLIC SIMULATOR

  • 所属行业:仪器仪表 气体检测分析仪 可燃性气体检测仪
  • 发布日期:2020-04-21
  • 阅读量:435
  • 价格:面议
  • 产品规格:不限
  • 产品数量:1.00 台
  • 包装说明:不限
  • 发货地址:上海闵行  
  • 关键词:VACUMED

    VACUMED 代谢模拟器METABOLIC SIMULATOR详细内容

    VACUMED  代谢模拟器METABOLIC SIMULATOR WITH MASS FLOW CONTROLLER
    The Ultimate Laboratory StandardDetail of Set Point Controller
    Why Metabolic Calibration?
    Ideally, the respiratory and metabolic variables (TV, RR, VE, VO2, VCO2) generated by the simulator should be indistinguishable from those produced by humans. The simulator produces and controls those variables very precisely. Consequently, one can test any breath-by-breath or mixing chamber metabolic measurement system with high accuracy and full confidence.
    Metabolic Calibration uses the principle of mass preservation to create known VO2 and VCO2. The principle of mass preservation dictates that the chosen mass flow of CO2 and O2 remains unchanged regardless of the ventilatory dilution (i.e. the minute ventilation that delivers it into the system under calibration).
    This top of the line metabolic calibration gas titration device consists of a Mass Flow Controller capable of delivering any set point flow of this gas (~21%CO2, 79%N2) within a range of 0 - 20 liters/min STPD and a Set Point Module (SPM) which enables selection of desired mass flows.
    Four switch-selected cal gas flow values of approx. 4.85, 9.70, 14.55 and 19.40 correspond to approx. 1, 2, 3 and 4 liters per minute of VO2 and VCO2. (Model 17057 has addditional 5 and 6 liter settings) The exact value will depend on physical parameters, such as barometric pressure, air temperature and relative humidity. The provided O2CAL.xls spreadsheet software enable precise determination of the selected metabolic rate and the percentage error of the metabolic measurement system being calibrated.
    The remaining two positions of the SPM's rotary switch correspond to a Zero flow setting and a continuous 10-turn potentiometer-operated selection of any desired mass flow within the range of 0 - 20 L/min STPD.
    Metabolic Calibration accuracy is +/-1% VO2 & VCO2.
    Most flow sensors are typically calibrated with a fixed volume manual syringe, but little is known about the performance of that flow sensor over the whole physiological range.
    This device includes a mass flow controller to produce precise Metabolic Calibration gas mixtures to verify VO2 and VCO2. It provides highest accuracy and ease of adjustment.
    So how does this affect the accuracy of your ventilation measurements (and ultimately VO2) when you have calibrated the flow sensor at some unknown manual stroke speed, then use the sensor for low-flow pediatric or resting measurements, or high flow VO2max measurements?
    What this device can be used for:
    ?	Produce an exact, simulated VO2 and VCO2 to verify the overall accuracy of a metabolic measurement system and its software.
    ?	Check the accuracy of any flow / volume measurement device.
    ?	Simulate breathing at rates from 6 to 80 breaths per minute.
    ?	Check the effects of various gas mixtures on a flow sensor.
    Specifications
    Model 17056 Stroke Volume Accuracy: +/-1/4%
    Model 17057 Stroke Volume Accuracy between 600cc to 6000cc = +/- 1/2% 
    between 300cc and 600cc = +/- 1 to 2 % 
    Not recommended below 300cc
    Motor speed is continuously variable from approx. 6 to 80 strokes per minute.
    Digital Display: Breaths per minute 
    Mass and volumetric flow of calibration gas 
    Barometric Pressure 
    Temperature
    Stroke Volume (Preset stops) 
    0.5, 1.0, 1.5, 2.0, 2.5, and 3.0 liters 
    (Other volume settings available on special order)
    #17056 Size: 12" (31cm) wide x 16" (41cm) high x 29" (74cm) long
    Voltage: 110 / 220 VAC (Specify)
    Weight: Approx 40 pounds (18 kg)
    A Metabolic Calibration gas containing 20.9% CO2, balance N2, is required.
    Included is an Excel file that allows you to make the required BTPS/STPD corrections automatically and print a calibration report.
    Other possible uses of this device:
    ?	This variable speed motor-driven calibration syringe will check the accuracy and repeatability of any flow sensor over the range from 3 to 240 L/min of VE. 
     
    ?	Produce any mixture of two gases and 
    change that mixture on-the-fly. 
    (Dilution formula available from VacuMed) 
     
    ?	Generate custom low-O2 mixtures for 
    simulated high altitude training. 
     
    ?	Quality Control of flow/volume sensing devices
    VACUMED  代谢模拟器METABOLIC SIMULATOR
    

    http://micforce.cn.b2b168.com
    欢迎来到米力光国际贸易有限公司网站, 具体地址是上海市闵行区上海市闵行区江协路52弄24号,联系人是周先生。 主要经营口腔生物材料试验机、口腔动静态疲劳试验机、齿科冷热疲劳循环仪、非接触式引伸计、光学引伸计、视频引伸计、光学应变引伸计、半导体热阻分析仪、稳态热阻分析仪、瞬态热阻分析仪、材料热阻测试仪、材料导热系数测试仪、可旋转动态热机械分析仪DMA DMTA 、高力值动态热机械分析仪、拉扭疲劳试验机、三轴疲劳试验机、双向拉压疲劳试验机、动静态疲劳试验机、拉金线、推金球、推锡球、推焊点、拔金球、拔焊点、拔锡球、。 单位注册资金单位注册资金人民币 100 万元以下。 本司供应microfet2,jamar,microntracker等产品,真正的为客户创造价值、节省费用,实现“设备永续、设备**”!